Vacuum Gauge Controller AGC-100
66
Transmit:
OFS
[,a,sx.xxxxEsxx] <CR>[<LF>]
Description
a
Mode, a =
0 –> Off (default)
No offset value needs to be entered
1 –> On
If no offset value has been entered, the pre-
viously defined offset value is taken over
2 –> Auto (offset measurement)
No offset value needs to be entered
3 –> Zero adjustment CDG
No offset value needs to be entered
sx.xxxxEsxx Offset
1)
, [in current pressure unit]
(default = 0.0000E+00)
s = sign
1)
Values can be entered in any format. They are internally converted
into the floating point format.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,sx.xxxxEsxx
<CR><LF>
Description
a Mode
sx.xxxxEsxx Offset
1)
, [in current pressure unit]
s = sign
Transmit:
SC1
[,a,b,c.ccEscc,d.ddEsdd] <CR>[<LF>]
Description
a
Gauge activation, a =
0 –> Manual (default)
1 –> Hot start
3 –> Via measurement channel 1
b
Gauge deactivation, b =
0 –> Manual (default)
1 –> Self control
3 –> Via measurement channel 1
c.ccEscc
ON threshold (s = sign)
d.ddEsdd
OFF threshold (s = sign)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c.ccEscc,d.ddEsdd <CR><LF>
Description
a Gauge
activation
b Gauge
deactivation
c.ccEscc
ON threshold (s = sign)
d.ddEsdd
OFF threshold (s = sign)
OFS
- Offset correction
(linear gauges)
5.7 Gauge Control
SC1
- Gauge control
Summary of Contents for AGC-100
Page 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Page 3: ......
Page 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Page 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ...t qna96e1 ...