Vacuum Gauge Controller AGC-100
74
Transmit:
MAC
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: aa-aa-aa-aa-aa-aa
<CR><LF>
Description
aa-aa-aa-aa-aa-aa
Ethernet MAC address of the unit:
00-A0-41-0A-00-00 … 00-A0-41-0B-FF-FF
Transmit:
PNR
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a.aa
<CR><LF>
Description
a.aa
Firmware version, e.g. 1.00
Transmit:
RHR
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a
Run (operating) hours, e.g. 24 [hours]
Transmit:
TAD
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: aa.aaaa
<CR><LF>
Description
aa.aaaa A/D
converter
Measurement signal [0.0000 … 11.0000 V]
Transmit:
TAI
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit:
<ENQ>
starts the test (very brief)
Receive: a.aa
<CR><LF>
Description
a.aa Identification
[kOhm]
MAC
- Ethernet MAC address
PNR
- Firmware version
RHR
- Operating hours
TAD
- A/D converter test
TAI
– ID resistance test