General Power Meter Functions
2
N1911A/1912A P-Series Power Meters User’s Guide
41
Figure 2-26
Channel offset indicator
Setting Display Offsets
This gain or loss is applied to the measured power after any channel
offsets or mathematical functions have been included.
Offsets are entered in dB and the range of values is –100 dB to +100 dB.
A positive value compensates for a loss, and a negative value compensates
for a gain.
Procedure
Enter a display offset on the currently selected window:
1
Press
to display the
Measurement Setup
screen.
2
Select the window you wish to set the offset value on by pressing the
key. The currently selected window/measurement is
displayed.
3
Use the
and
keys to highlight the
Offset
setting field.
4
Press
to check the
Offset
setting field.
Offset Indicator
N O T E
The
Ofs
symbol is not displayed when the associated measurement is displayed in Dual
Numeric or Analog format.
Meas Select