General Power Meter Functions
42
N1911A/1912A P-Series Power Meters User’s Guide
Figure 2-27
Typical display offset display
5
Press
to highlight the
Offset
value field and press
to display
the
Display Offset
pop- up. Use the numeric keypad to enter the required
value in the
Offset
pop- up window.
6
Confirm your choice by pressing
. Press
key to complete
the offset entry.
Setting Frequency Dependent Offsets
Frequency dependent offset tables provide a quick and convenient method
of compensating for frequency related changes in the response of your test
system. Note that when selected, frequency dependent offset corrections
are applied IN ADDITION to any correction for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset
tables with a maximum of 80 frequency points each.
Measurement
Selected Window/
Offset Setting Field
Offset Value Field
dB
N O T E
The
Ofs
indicator is displayed if Display Offset is selected.