General Power Meter Functions
80
N1911A/1912A P-Series Power Meters User’s Guide
Frequency Sweep Mode
Frequency sweep is generally used in a frequency response calibration
system where the amplitude is fixed, and the frequency of the power
source signal is swept. This mode can be used to determine the frequency
response of a device under test.
Procedures
1
Connect sensor to a power source.
2
Connect power meter
TRIG OUT
to power source
TRIG IN
using a
BNC cable. The same connection is applied for power source
TRIG OUT
to power meter
TRIG IN
. See
Figure 2- 58
.
3
Press
. The
Channel Setup
displays as shown in
Figure 2-59
.
N O T E
This feature is only available when N8480 Series, E-Series E4410, E-Series E9300, E-Series
E9320 or P-Series N1920 sensor is connected.
N O T E
When N8480 Series, E-Series E4410 or E-Series E9300 sensor is connected,
Sensor Mode
is set to
AVG
only
by default.