Using E4410 E-Series Power Sensors
6
N1911A/1912A P-Series Power Meters User’s Guide
169
Power Meter Configuration
The P- Series power meters automatically recognize an E- Series E4410
power sensor when it is connected. The sensor calibration data is
automatically read by the power meter. Also, the power meter
automatically configures the averaging as shown in
Figure 6- 120
.
Figure 6-120
E-Series CW sensor auto-averaging settings
10 dB
10 dB
Maximum Sensor Power
Minimum Sensor Power
Power
Sens
or
Dy
n
a
mi
c
Ra
ng
e
10 dB
10 dB
1
2
3
4
1
1
1
8
Resolution Setting
Numb
er of
A
v
er
ages
1
1
1
16
1
1
2
32
1
1
16
256
1
8
128
128
N O T E
These values are valid only for the power meter channel connected with E-Series E4410
power sensor. Averaging settings can also be manually configured.