8
Using N8480 Series Power Sensors
200
N1911A/1912A P-Series Power Meters User’s Guide
N8480 Series Sensors Connection Requirements
Table 8-15
N8480 Series connection requirements
Sensor
Connection Requirements
N8481A
N8481H
N8482A
N8482H
These power sensors connect directly to the POWER REF.
N8481B
N8482B
These power sensors are configured with an attenuator. Prior to
calibration this attenuator must be removed. Replace the attenuator
before making measurements.
N8485A
This power sensor requires an APC 3.5 (f) to 50
Ω
(m) N-Type adapter
(08485-60005) to connect to the POWER REF.Remove this adapter before
making measurements.
N8486AR
N8486AQ
These waveguide power sensors have two connectors. Use the
N-Type connector to calibrate the power meter.
N8487A
N8488A
This sensor requires an APC 2.4 (f) to 50
Ω
(m) N-Type adapter
(08487-60001) to connect to the POWER REF. Remove this adapter
before making measurements.