General Power Meter Functions
46
N1911A/1912A P-Series Power Meters User’s Guide
To view the frequency dependent offset tables currently stored in the
power meter, press
, ,
.
The
Frequency
Dependent Offset Tables
screen is displayed as shown in
Figure 2- 28
.
Editing frequency dependent offset tables requires the following steps:
1
Identify and select the table you want to edit
2
Rename the table
3
Enter the frequency and offset pairs
4
Save the table
Procedure
First select the table you want to edit as follows:
1
Press
,
,
to display the
Offset Tbls
screen.
2
Choose the table you want to edit using the
and
keys. Press
to display the
Edit Offset
screen as shown in
Figure 2- 30
.
Figure 2-30
“Edit Offset” display with data added
Tables
Freq. Dep. Offset
N O T E
A frequency in the range of 0.001 MHz to 1000.0 GHz can be entered. A calibration factor in
the range of 1 % to 150 % can be entered.
Tables
Freq. Dep. Offset
Edit Table