Agilent InfinityLab LC Series Vialsamplers User Manual
99
Installation of the Optional Sample Cooler or Thermostat
4
Using the Sample Cooler or Thermostat in an Infinity II Sampler
Control
The
Thermostat
section of the
Control
dialog box is available when you have a
sample cooler or sample thermostat installed. It allows you to switch the
cooler or thermostat on and off manually.
Select
On
to switch on the cooler/thermostat. Specify the required
temperature in the adjacent field. Note that the specified temperature must be
at least 5 °C below ambient for proper temperature control.
Select
Off
to switch off the cooler/thermostat.