![Agilent Technologies 1260 Infinity II Series User Manual Download Page 38](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552038.webp)
38
Agilent InfinityLab LC Series Vialsamplers User Manual
2
Site Requirements and Specifications
Site Requirements
Room Size and Ventilation
WA R N I N G
Flammable refrigerant
Formation of flammable gas-air mixtures inside the sample thermostat and
laboratory.
➔
Keep open fire or sources of ignition away from the device.
➔
Ensure a room size of 4 m
3
(1 m
3
for every 8 g of R600a refrigerant inside of the
sample thermostat).
➔
Ensure adequate ventilation: typical air exchange of 25 m
3
/h per m
2
of laboratory
floor area.
➔
Keep all ventilation openings in the enclosure clear of obstructions. Do not block
the openings on the circumference of the sample thermostat.