Agilent InfinityLab LC Series Vialsamplers User Manual
9
Maintenance and Repair
Exchange the Gripper Arm
Exchange the Gripper Arm
When
Defective gripper arm
Tools required
Description
Straightened paper clip.
Parts required
p/n
Description
G1313-60010
Gripper assembly
Preparations
Finish any pending acquisition job
1
In the Local Controller start the maintenance mode and
select
Change Gripper
function.
OR
In the
Tools
section of the Agilent Lab Advisor software
select
Service & Diagnostics > Maintenance Positions
> Change Gripper
, click
Start
and wait until the gripper
is in maintenance position.
2
Open the front door.
3
Identify the slit below the gripper motor and the gripper
arm release button.