Agilent InfinityLab LC Series Vialsamplers User Manual
3
Using the Module
Install the Column Shelf
Install the Column Shelf
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Preparations
•
Switch off the pump
•
Switch off the sampler
•
Open the doors of the sampler
•
Open the ICC
WA R N I N G
Fire or explosion due to missing leak drainage
Solvents may run into the module if Column Shelf or Integrated Column
Compartment are not installed.
➔
Install either Column Shelf or Integrated Column Compartment before starting the
module.
1
Remove the capillary connection from the integrated
column compartment to the module.
2
Remove the column.