52
Agilent InfinityLab LC Series Vialsamplers User Manual
2
Site Requirements and Specifications
Specifications of the Integrated Column Compartment
Performance Specifications Agilent InfinityLab LC Series
Integrated Column Compartment (G7130A)
Table 10
Performance Specifications Agilent InfinityLab LC Series Integrated Column
Compartment (G7130A)
Type
Specification
Comments
Temperature range
5 °C above ambient to 80 °C
Column capacity
2 columns up to 30 cm and
4.6 mm ID
Temperature stability
±
0.10 °C
at sensor
Temperature accuracy
±
0.8 K (
±
0.5 K with calibration)
measured with water at
ambient temperature (25 °C),
set point at 40 °C and a stable
flow of 1 mL/min at outlet of
pre-heater
Warm up time
20 – 40 °C in 5 min
at sensor