Agilent InfinityLab LC Series Vialsamplers User Manual
199
Maintenance and Repair
9
Exchange the Needle Seat Assembly
6
In Lab Advisor use
Next
to move the needle in the down
position.
The needle will move slowly downwards and will stop
approximately 2 mm above the seat.
7
Disconnect the sample loop from the needle.
8
Loosen the fixing screw.
9
Lift out the needle.
10
Disconnect the seat-capillary fitting from the injection
valve (port 5).
11
Use a small flat-head screwdriver to ease out the needle
seat.