Agilent InfinityLab LC Series Vialsamplers User Manual
215
Maintenance and Repair
9
Exchange the Metering Seal and Piston
14
Install the metering head assembly in the autosampler.
Ensure the large hole in the metering head is facing
downwards. Tighten the bolts alternately two turns at a
time until the metering device is secure.
15
Reinstall the capillaries.
16
In the Local Controller exit the maintenance mode and
select
Change metering device
function.
OR
In the
Tools
section of the Agilent Lab Advisor software
system screen exit
Service & Diagnostics >
Maintenance Positions > Change Metering Device
:
click
Next
and wait until the metering device is in Home
position.
17
Select
Back
to finish the procedure.
18
Close the doors.
19
Perform a pressure test.