204
Agilent InfinityLab LC Series Vialsamplers User Manual
9
Maintenance and Repair
Exchange the Sample Loop Assembly
6
Disconnect the loop capillary from the metering device
and from the needle.
7
Install the new sample loop.
N O T E
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
8
After installing the new sample loop, check the needle
tip position by using the
Needle Control
function in Lab
Advisor.
9
Click on
Next
in the LabAdvisor Needle lift up again.
10
Turn or flap the wash port to the right.
11
Install the safety cover (1.) and fix the screw (2.).