![Agilent Technologies 1260 Infinity II Series User Manual Download Page 292](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552292.webp)
292
Agilent InfinityLab LC Series Vialsamplers User Manual
13
LAN Configuration
Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the
non-volatile memory of the module. It never affects the currently active
parameters. Therefore, manual configuration can be done at any time. A
power cycle is mandatory to make the stored parameters become the active
parameters, given that the initialization mode selection switches are allowing
it.
Figure 50
Manual Configuration (Principle)
7(/1(7
6HVVLRQ
&RQWURO
0RGXOH
6WRUHG
3DUDPHWHU
1RQ9RODWLOH
5$0