Agilent InfinityLab LC Series Vialsamplers User Manual
3
Using the Module
Install the External Tray
Install the External Tray
Parts required
p/n
Description
G7129-60000
External Tray for 5 x 2 mL Vials
G1313-27302
Disposal tube
Preparations
•
Turn the sampler off
•
Open the doors
•
Remove the samples
N O T E
Keep foam and plastic cover in a safe place.
For best temperature performance, and if the external tray is not in use, it is best to cover
the opening for the external tray with the original parts.
1
Lift the foam with a flathead screwdriver.
2
Remove the foam.