194
Agilent InfinityLab LC Series Vialsamplers User Manual
9
Maintenance and Repair
Exchange the Needle Assembly
6
In Lab Advisor use
Next
to move the needle in the down
position.
The needle will move slowly downwards and will stop
approximately 2 mm above the seat.
7
Disconnect the sample loop from the needle.
8
Loosen the fixing screw.
9
Lift out the needle.
10
Insert the needle (1). Align the needle in the seat, then
tighten the screw firmly (2).
11
Align the needle in the seat (needle may be bent slightly
manually, if not aligned properly).