Agilent 1260 Infinity DAD and MWD User Manual
81
Using the Detector
4
Special Settings of the Detector
Run Recovery Settings
The detector supports run buffering, which means that an amount of run data
(*.uv and *.ch files) is stored in a storage medium (CompactFlash Card) in the
detector until either it is overwritten or the detector undergoes a power cycle.
If there is a temporary network failure or the PC is not able to constantly take
the data, the stored data is transferred to the ChemStation automatically
when the network connection is restored or the PC can take the data, so that
no loss of data occurs.
If there is a permanent network failure, the Run Recovery dialog box allows
you to restore the stored data to the data directory. From there you can copy
the files to the directory where the files are corrupted or not complete.
C A U T I O N
For this recovery mode the CompactFlash Card must be in the detector.
If the LAN communiaction is interruppted, no data is stored.
➔
Have the CompactFlash Card always inserted.
N O T E
On very large recovery files it may take a long time to restore it to the Agilent
ChemStation.
A sequence will be stopped in case of a network problem.
N O T E
When during recovery an error “Method/Sequence stopped” appears, the instrument
logbook shows an entry “No Run data available in device“.
In this case refer to
Summary of Contents for 1260 Infinity G1315C
Page 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Page 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Page 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Page 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Page 313: ......