Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
When
If the capillary is blocked
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Wrench, 4 mm
(for capillary connections)
Screwdriver, Pozidriv #1 PT3
Parts required
Description
For parts see
Preparations
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell,
N O T E
All descriptions in this procedure are based on the default orientation of the cell (as it is
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
N O T E
The fittings at the flow cell body are special types for low dead volumes and not
compatible with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.
Summary of Contents for 1260 Infinity G1315C
Page 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Page 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Page 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Page 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Page 313: ......