![Agilent Technologies 1260 Infinity G1315C User Manual Download Page 278](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-g1315c/1260-infinity-g1315c_user-manual_2867535278.webp)
278
Agilent 1260 Infinity DAD and MWD User Manual
13
LAN Configuration
Automatic Configuration with BootP
Assigning IP Addresses Using the Agilent BootP Service
The Agilent BootP Service assigns the Hardware MAC address of the
instrument to an IP address.
Determining the MAC address of the instrument using BootP Service
1
Power cycle the Instrument.
2
After the instrument completes self-test, open the log file of the BootP
Service using Notepad.
•
The default location for the logfile is
C:\Documents and Settings\All Users\
Application Data\Agilent\BootP\LogFile
.
•
The logfile will not be updated if it is open.
The contents will be similar to the following:
3
Record the hardware (MAC) address (for example, 0010835675AC).
4
The Error means the MAC address has not been assigned an IP address and
the Tab File does not have this entry. The MAC address is saved to the Tab
File when an IP address is assigned.
5
Close the log file before turning on another instrument.
6
Uncheck the
Do you want to log BootP requests?
box after configuring
instruments to avoid having the logfile use up excessive disk space.
Adding each instrument to the network using BootP
1
Follow
Start > All Programs > Agilent BootP Service
and select
Edit BootP
Settings
. The BootP Settings screen appears.
2
Uncheck the
Do you want to log BootP requests?
once all instruments have been
added.
02/25/10 15:30:49 PM
Status: BootP Request received at outermost layer
Status: BootP Request received from hardware address: 0010835675AC
Error: Hardware address not found in BootPTAB: 0010835675AC
Status: BootP Request finished processing at outermost layer
Summary of Contents for 1260 Infinity G1315C
Page 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Page 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Page 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Page 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Page 313: ......