Agilent 1260 Infinity DAD and MWD User Manual
181
Maintenance
9
Maintenance of High Pressure Flow Cell
1
Use a 4 mm hex key to unscrew the window assembly
[1] and remove the gasket [2] from the cell body.
If you want to replace the gasket only, continue with
step 8,
“Maintenance of Standard, Semi-Micro or Micro
2
Use a tooth pick to remove the quartz window from the
window assembly.
N O T E
If the washers fall out of the window assembly,
they must be inserted in the correct order with the
PTFE ring to prevent any leaks from the flow cell
window.
3
“Maintenance of Standard, Semi-Micro or Micro Flow Cell”
on page 176 for reassembling.