Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Replacing Capillaries on a Semi-Micro and Micro Flow Cell
5
Use a small flat screw driver to carefully lift off the I.D.
tag. Shown is the default orientation. See Note at the
beginning of this section.
6
Carefully insert the I.D. tag into the new heat exchanger.
Shown is the default orientation. See Note at the
beginning of this section.
7
Fix the new heat exchanger to the clamp unit and the
heat exchanger to the cell body.
8
Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.