Agilent 1260 Infinity DAD and MWD User Manual
283
LAN Configuration
13
Storing the settings permanently with Bootp
Storing the settings permanently with Bootp
If you want to change parameters of the module using the Bootp follow the
instructions below.
1
Turn off the module.
2
Change the module's settings of the Configuration Switch to
“Bootp &
Store”
3
Start the Agilent Bootp Service and open its window.
4
If required, modify the parameters for the module according to your needs
using the existing configuration.
5
Press
OK
to exit the Bootp Manager.
6
Now turn on the module and view the Bootp Server window. After some
time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the module.
7
Close the Agilent Bootp Service and turn off the module.
8
Change the settings of the module’s Configuration Switch to
“Using Stored”
mode, see
9
Power cycle the module. The module can be accessed now via LAN without
the Agilent Bootp Service.
Summary of Contents for 1260 Infinity G1315C
Page 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Page 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Page 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Page 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Page 313: ......