Agilent 1260 Infinity DAD and MWD User Manual
33
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications G1365D
Table 6
Performance Specifications G1365D
Type
Specification
Comments
Detection type
1024-element photodiode array
Light source
Deuterium and tungsten lamps
The UV-lamp is equipped with
RFID tag that holds lamp typical
information.
Data rate
up to 20 Hz
Wavelength range
190 – 950 nm
Short term noise
(ASTM) Single and
Multi-Wavelength
< ± 0.7·10
-5
AU at 254 and 750 nm
see
"Specification Conditions"
below
Drift
< 0.9·10
-3
AU/h at 254 nm
see
"Specification Conditions"
below
Linear absorbance
range
> 2 AU (5 %) at 265 nm
see
"Specification Conditions"
below
Wavelength
accuracy
± 1 nm
Self-calibration with deuterium
lines, verification with holmium
oxide filter
Wavelength
bunching
1 – 400 nm
Programmable in steps of 1 nm
Slit width
1, 2, 4 , 8, 16 nm
Programmable slit
Diode width
< 1 nm