Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Correcting Leaks
Correcting Leaks
1
Use tissue to dry the leak sensor area and the leak pan.
2
Observe the capillary connections and the flow cell area for leaks and
correct, if required.
Figure 61
Observing for Leaks
3
Replace the front cover.
When
If a leakage has occurred in the flow cell area or at the heat exchanger or at the capillary
connections
Tools required
p/n
Description
Tissue
Wrench, 1/4 inch
for capillary connections
5043-0915
Fitting mounting tool
for bio-inert capillaries
Preparations
Remove the front cover.
/HDNSDQ
/HDNVHQVRU