Agilent 1260 Infinity DAD and MWD User Manual
4
Using the Detector
Setting up an Analysis
5
Purge the pump. For more information see
6
Allow the detector to warm up at least 60 minutes to provide a stable
baseline (see example in
Figure 17
Stabilization of Baseline (both lamps turned on at the same time)
N O T E
For reproducible chromatography, the detector and lamp should be on for at least one hour.
Otherwise the detector baseline may still drift (depending on the environment). See also
section
Wander/Drift Problems Due to Temperature Changes
in the
Service Manual.
.
Table 11
Baseline drift after lamp turn on (example from Figure above)
Time [minutes]
Drift [mAU/hr]
17 - 20
2.6
27 - 30
0.8
37 - 40
0.4
47 - 50
0.2
57 - 60
< 0.2