116
Agilent 1260 Infinity DAD and MWD User Manual
5
How to optimize the Detector
Optimizing Selectivity
Figure 43
Selectivity by Ratio Qualifiers
In a four-component mixture, only biphenyl was recorded. The other three
peaks were suppressed because they did not meet the ratio-qualifier criterion
and therefore the output was set to zero. The characteristic wavelengths
249 nm (
λ
1
) and 224 nm (
λ
2
) were found from the spectra shown in
on page 115. The ratio range was set at 2 – 2.4 (2.2 ±10%). Only when the ratio
between 249 and 224 nm was within this range, is the signal plotted. Of all
four peaks, only the third fulfilled the criterion (
others were not plotted.
6LJQDOVDWQP
1RVHOHFWLYLW\
%LSKHQ\O
R7HUSKHQ\O
7LPHPLQ
:LWKUDWLRTXDOLILHU
QP