Agilent 1260 Infinity DAD and MWD User Manual
137
Error Information
7
Detector Error Messages
UV Heater Current
Error ID: 7453
The UV lamp heater current is missing.
During UV lamp ignition, the processor monitors the heater current. If the
current does not rise above the lower limit within one second, the error
message is generated.
Calibration Values Invalid
Error ID: 1036
The calibration values read from the spectrometer ROM are invalid.
After recalibration, the calibration values are stored in ROM. The processor
periodically checks if the calibration data are valid. If the data are invalid or
cannot be read from the spectrometer ROM, the error message is generated.
Probable cause
Suggested actions
1
Lamp disconnected.
Ensure the UV lamp is connected.
2
Ignition started without the top foam in
place.
Please contact your Agilent service
representative.
3
Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
4
Defective detector main board.
Please contact your Agilent service
representative.
5
Defective power supply.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Defective connector or cable.
Please contact your Agilent service
representative.
2
Defective PDA/optical unit.
Please contact your Agilent service
representative.
Summary of Contents for 1260 Infinity G1315C
Page 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Page 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Page 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Page 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Page 313: ......