The calculated ‘Operating levels’ should respect the following tolerances:
"
The tolerances may be different according to local regulations.
10
These charts will be used for the daily quality test. For more information, refer
to
‘Performing the daily QC test’
.
Operating level
Tolerance
Base + Fog
- 0.03 / + 0.05
Mid density
+/- 0.20
DD (Density difference)
+/- 0.20
MaxD
+/- 0.25