Verifying Acceptable Spatial Resolution and Artifact Levels
To establish the image geometry reference values, proceed as follows:
1
Print the 'QC ' test image or use the previously printed QC test image used to
establish the daily operating density levels.
You should obtain an image looking like this:
Drystar 5300 Quality Control Test Image
Date : __________
Initials : _____
Sharpness Test
Geometry Test
Density test
Dimension A
Dimension B
Each cluster
has 3 groups
of dots of
different sizes
Each group has 5 dots
3 Resolution clusters
MaxD (100%):
DD (Density Difference):
(High-Low)
Mid Density:
Base + Fog: