105
2920B EN 20031206
4
When the film calibration has been completed successfully, three cases are
possible:
•
Calibration successful, results OK.
The obtained MaxD (3.10) after calibration, is displayed:
(xxxxx) is the code of the Densitometer type (refer to
‘Changing the default densitometer’
on page
•
Calibration successful, but MaxD differs more than 2% from the target value.
The obtained MaxD (3.10) after calibration, is displayed:
(xxxxx) is the code of the Densitometer type (refer to
‘Changing the default densitometer’
on page
•
Calibration successful, but the maximum density has been reached.
The obtained MaxD (3.10) after calibration, is displayed:
(xxxxx) is the code of the Densitometer type (refer to
‘Changing the default densitometer’
on page
5
Press the Confirm key to return to the Calibration menu.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise the old calibration parameters
will be used.
Film
calibration OK
MaxD = 3.10 (xxxxx)
CA
Flashing
Flashing
Retry
calibration or
Resume printing
MaxD = 3.10 (xxxxx)
CA
Flashing
Flashing
Film
calibration
MaxD is reached
MaxD = 3.10 (xxxxx)
CA
Flashing
Flashing
Summary of Contents for Drystar 5300
Page 1: ...Drystar 5300 Reference manual POWER ...
Page 160: ...160 2920B EN 20031206 ...
Page 168: ...168 2920B EN 20031206 ...
Page 189: ...Equipment information sheet Appendix A ...
Page 195: ...Drystar media density response data Appendix B ...
Page 198: ...198 2920B EN 20031206 ...
Page 199: ...Quality Control Charts Appendix C ...
Page 205: ...Index Appendix D ...
Page 210: ...Printed in Belgium Published by Agfa Gevaert N V B 2640 Mortsel Belgium 2920B EN 20031206 ...