2920B EN 20031206
Changing the default densitometer
1
Perform steps 1 and 2 of the
‘Accessing the Change settings menu’
procedure on page
2
On the ‘Change settings’ screen, press the Down key three times, followed by
the Confirm key to select ‘Image quality’.
The following screen will appear:
3
Press the Down key, followed by the Confirm key to select
‘Def. densitometer’.
The ‘Select default densitometer’ screen is displayed:
The current default densitometer is indicated with a ‘*’.
4
Press the up/down key to select the default densitometer.
"
All densities that are shown on the Drystar 5300 screens are according to
the response of the default densitometer. Refer also to
on page
5
Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
CS
1 General setting
2 Def. densitometer
3 Modality16char
4 Modality16char
5 etc...
CS
Select default
densitometer:
1 * MacBeth TR924
2 X-Rite 310
3 X-Rite 331
4 X-Rite 341
5 X-Rite 301