The TSP head, particle trap and cyclone disassemble easily for cleaning.
Figure 6-20 The TSP head and cyclone can be disassembled for cleaning.
Replace pumps in pump module
The pumps in the Particle Monitor pump module and in the Particle Profiler pump module can easily
be replaced.
The Particle Monitor contains a sample pump: (
AQS R10
) and a purge pump (
AQS R34
). The Particle
Profiler only has a sample pump (
AQS R10
)
Typically the pumps last between 12 and 18 months. However the pumps should be replaced when
the correct flow can no longer be maintained.
Figure 6-21 The Particle Monitor contains two pumps, the Particle Profiler has one pump.
Leak check gas module
If the correct flow rate cannot be achieved through a gas module, this may indicate a leak. An
individual gas module can be checked for leaks by using a small diaphragm pump and a flow meter in
the configuration shown in Figure 6-22
Tools required
Diaphragm pump R10, flowmeter DS R8 or AQM R7
Parts required
None
Service interval
Only as required. (See Section 6.1)
Online reference
None
Online video
None