Data Format and Input Range
•
91
Input Range
Resolution
±
15 mV,
±
50 mV
1
µ
V
three decimal places
±
100 mV,
±
150 mV,
±
500 mV
10
µ
V
two decimal places
±
1 V,
±
2.50 V,
±
5 V
100
µ
V
four decimal places
±
10 V
1mV
three decimal places
±
20 mA
1
µ
A
three decimal places
Type J and T thermocouple, RTD
0.01
°
C
two decimal places
Type K, E, R, S, B, N and C
thermocouple
0.1
°
C
one decimal places
Table 4 -1 Data format and resolution
Example 1:
•
Input Range is
±
5 V
•
Input is -1.37 Volts
engineering units:
-1.3700<CR>
Example 2:
•
Input Range is
±
10 V
•
Input is +3.653 Volts
engineering units:
+03.653<CR>
Example 3:
•
Input Range is Type K thermocouple (range 0
°
C to 1000
°
C)
•
Input is 406.5
°
C
engineering units:
+0406.5<CR>
Example 4:
•
Input Range is Type T therm ocouple (range -100
°
C to 400
°
C)
•
Input is -50.5
°
C
engineering units:
-
050.50<CR>
Percent of FSR (Full Scale Range)
•
Data format bit 1 and 0 set to 01 is percent of FSR.
•
This data format including three components.
1. sign (+ or - )
2. digits
3. decimal point