104
•
Calibration
Table 5-2:
ND-6013 Calibration Resistance
Code
Input Range
Span Calibration
Resistance
Offset Calibration
Resistance
20
Pt-100, -100
°
C to +100
°
C,
α
=.00385
200
Ω
50
Ω
21
Pt-100, 0
°
C to +100
°
C,
α
=.00385
200
Ω
50
Ω
22
Pt-100, 0
°
C to +200
°
C,
α
=.00385
200
Ω
50
Ω
23
Pt-100, 0
°
C to +600
°
C,
α
=.00385
350
Ω
50
Ω
24
Pt-100, -100
°
C to +100
°
C,
α
=.003916
200
Ω
50
Ω
25
Pt-100, 0
°
C to +100
°
C,
α
=.003916
200
Ω
50
Ω
26
Pt-100, 0
°
C to +200
°
C,
α
=.003916
200
Ω
50
Ω
27
Pt-100, 0
°
C to +600
°
C,
α
=.003916
350
Ω
50
Ω
28
Ni-100, 0
°
C to +100
°
C
200
Ω
50
Ω
29
Ni-120, 0
°
C to +100
°
C
350
Ω
50
Ω
2A
0 Ohm to 60Ohm
60
Ω
0
Ω