50
•
Command Set
3.12 Span Calibration to each Channel
@Description
To correct the gain errors of AD converter by using the span calibration.
@Syntax
$(Addr)0(Channel No)<CR>
$
Command leading code
(1 character)
(Addr)
Address ID
(2 character)
0
Span calibration
(1 character)
(Channel No)
Channel for Calibration
(1 character) 0~2
@Response
!(Addr)<CR>
or
?(Addr)<CR
>
!
Command is valid.
?
Command is invalid.
(Addr)
Address ID.
@Example
User command:
$0601<CR>
Response:
!06<CR>
To perform the span calibration for analog input module ND-6013
channel 1 , address ID is 06H.
Note:
To perform the calibration, a proper input signal should be
connected to the analog input module. Different input range
have different input voltage, detail refer chapter 5
“
Calibration
” .
(for 6013 F/W version
A3.05~A4.60)
Summary of Contents for NuDAM-6017