Calibration
•
103
Analog Input Module‘s Calibration Voltages
Table 5-1: ND-6011/D/ND-6018 Calibration voltages
Code
Input Range
Offset Calibration
voltage
Span Calibration
Voltage
00
±
15 mV
0 mV
+15 mV
01
±
50 mV
0 mV
+50 mV
02
±
100 mV
0 mV
+100 mV
03
±
500 mV
0 mV
+500 mV
04
±
1 V
0 V
+1 V
05
±
2.5 V
0 V
+2.5 V
06
±
20 mA
0 mA
+20 mA
0E
J
(0
°
C to 760
°
C)
0 mV
+50 mV
0F
K
(0
°
C to 1000
°
C)
0 mV
+55 mV
10
T
(-100
°
C to 400
°
C)
0 mV
+30 mV
11
E
(0
°
C to 1000
°
C)
0 mV
+78 mV
12
R
(500
°
C to 1750
°
C)
0 mV
+25 mV
13
S
(500
°
C to 1750
°
C)
0 mV
+19.5 mV
14
B
(500
°
C to 1800
°
C)
0 mV
+15 mV
15
N
(-270
°
C to 1300
°
C)
0 mV
+55 mV
16
C
(0
°
C to 2320
°
C)
0 mV
+39 mV
Table 5-1: ND-6012/D/ND-6017 Calibration voltages
Code
Input Range
Offset Calibration
voltage
Span Calibration
Voltage
08
±
10 V
0 mV
+10 V
09
±
5 V
0 mV
+5 V
0A
±
1 V
0 mV
+1 V
0B
±
500 mV
0 mV
+500 mV
0C
±
150 mV
0 mV
+150 mV
0D
±
20 mA
0 mA
+20 mA