Data Format and Input Range
•
97
The following table shows the relation between the input range setting with the
data format and the resolution when using ND-6011/D or 6018 to measure
temperature by thermocouple.
Code
Input Range
Thermocouple
Data
Format
Maximum
Minimum
Displayed
Resolution
0E
J
(0
°
C to 760
°
C)
Eng. Units
+760.00
+000.00
0.01
°
C
0F
K
(0
°
C to 1000
°
C)
Eng. Units
+1000.0
+0000.0
0.1
°
C
10
T
(-100
°
Cto400
°
C)
Eng. Units
+400.00
-100.00
0.01
°
C
11
E
(0
°
C to 1000
°
C)
Eng. Units
+1000.0
+0000.0
0.1
°
C
12
R
(500
°
C to 1750
°
C)
Eng. Units
+1750.0
+0500.0
0.1
°
C
13
S
(500
°
C to 1750
°
C)
Eng. Units
+1750.0
+0500.0
0.1
°
C
14
B
(500
°
C to 1800
°
C)
Eng. Units
+1800.0
+0500.0
0.1
°
C
15
N
(-270
°
C to 1300
°
C)
Eng. Units
+1300.0
-0270.0
0.1
°
C
16
C
(0
°
C to 2320
°
C)
Eng. Units
+2320.0
+0000.0
0.1
°
C
Code
Input Range
Thermocouple
Data
Format
Maximum
Minimum
Displayed
Resolution
0E
J
(0
°
C to 760
°
C)
% of FSR
+100.00
+000.00
0.01%
0F
K
(0
°
C to 1000
°
C)
% of FSR
+100.00
+000.00
0.01%
10
T
(-100
°
Cto400
°
C)
% of FSR
+100.00
-025.00
0.01%
11
E
(0
°
C to 1000
°
C)
% of FSR
+100.00
+000.00
0.01%
12
R
(500
°
C to 1750
°
C)
% of FSR
+100.00
+028.57
0.01%
13
S
(500
°
C to 1750
°
C)
% of FSR
+100.00
+028.57
0.01%
14
B
(500
°
C to 1800
°
C)
% of FSR
+100.00
+027.27
0.01%
15
N
(-270
°
C to 1300
°
C)
% of FSR
+100.00
-020.76
0.01%
16
C
(0
°
C to 2320
°
C)
% of FSR
+100.00
+000.00
0.01%
Code
Input Range
Thermocouple
Data
Format
Maximum
Minimum
Displayed
Resolution
0E
J
(0
°
C to 760
°
C)
2’s Comp.
7FFF
0000
1 LSB
0F
K
(0
°
C to 1000
°
C)
2’s Comp.
7FFF
0000
1 LSB
10
T
(-100
°
Cto400
°
C)
2’s Comp.
7FFF
E000
1 LSB
11
E
(0
°
C to 1000
°
C)
2’s Comp.
7FFF
0000
1 LSB
12
R
(500
°
C to 1750
°
C)
2’s Comp.
7FFF
2492
1 LSB
13
S
(500
°
C to 1750
°
C)
2’s Comp.
7FFF
2492
1 LSB
14
B
(500
°
C to 1800
°
C)
2’s Comp.
7FFF
238E
1 LSB
15
N
(-270
°
C to 1300
°
C)
2’s Comp.
7FFF
E56B
1 LSB
16
C
(0
°
C to 2320
°
C)
2’s Comp.
7FFF
0000
1 LSB
Summary of Contents for NuDAM-6017