Calibration
•
99
2.
Apply the correct
offset resistance
to the analog input module
ND-6013 channel 0, detail resistance value, see table 5 -2
3.
Send “
Offset Calibration
$(Addr)10
” to analog input module
ND-6013 channel 0 .
4.
Apply the correct
span resistance
to the analog input module
ND-6013 channel 0, detail resistance value, see table 5 -2
5.
Send “
Span Calibration
$(Addr)00
” to analog input module
ND-6013 channel 0 .
6.
Apply the correct
offset resistance
to the analog input module
ND-6013 channel 1, detail resis tance value, see table 5 -2
7.
Send “
Offset Calibration
$(Addr)11
” to analog input module
ND-6013 channel 1 .
8.
Apply the correct
span resistance
to the analog input module
ND-6013 channel 1, detail resistance value, see table 5 -2.
9.
Send “
Span Calibration
$(Addr)01
” to analog input module
ND-6013 channel 1 .
10. Apply the correct
offset resistance
to the analog input module
ND-6013 channel 2, detail resistance value, see table 5 -2.
11. Send “
Offset Calibration
$(Addr)12
” to analog input module
ND-6013 channel 2 .
12. Apply the correct
span resistance
to the analog input module
ND-6013 channel 2, detail resistance value, see table 5 -2.
13. Send “
Span Calibration
$(Addr)02
” to analog input module
ND-6013 channel 2 five times.
14. Repeat procedure 2 to procedure 13 two times.
Calibration Procedure for ND-6013 Firmware Rev C4.60
1. Only open channel one and disable the other two channels.
2. Disable open detect function
3. Select the correct input range, different input range have different
apply calibration resistances.
4. Apply the correct
offset resistance
to
channel
0
, detail resistance
value, see table 5 -2.
5. Send “
Offset Calibration
$(Addr)1
” to analog input module five
times.
6. Apply the correct
span resistance
to
channel 0
, detail resistance
value, see table 5 -1
7. Send “
Span Calibration
$(Addr)0
” to analog input module
five times..
Repeat procedure 3 to procedure 7 two times