94
•
Data Format and Input Range
4.2 Analog Input Range
The following table shows the relation between the input range setting
with the data format and the resolution.
Engineering Units Table:
Code
Input
Range
Data
Format
+Full
Scale
Zero
- Full
Scale
Displayed
Resolution
00
±
15mV
Eng. Units
+15.000
±
00.000
-15.000
1
µ
V
01
±
50mV
Eng. Units
+50.000
±
00.000
-50.000
1
µ
V
02
±
100mV
Eng. Units
+100.00
±
000.00
-100.00
10
µ
V
03
±
500mV
Eng. Units
+500.00
±
000.00
-500.00
10
µ
V
04
±
1V
Eng. Units
+1.0000
±
0.0000
-1.0000
100.00
µ
V
05
±
2.5V
Eng. Units
+2.5000
±
0.0000
-2.5000
100.00
µ
V
06
±
20mA
Eng. Units
+20.000
±
00.000
-20.000
1
µ
A
07
Reserved
08
±
10V
Eng. Units
+10.000
±
00.000
-10.000
1mV
09
±
5V
Eng. Units
+5.0000
±
0.0000
-5.0000
100.00
µ
V
0A
±
1V
Eng. Units
+1.0000
±
0.0000
-1.0000
100.00
µ
V
0B
±
500mV
Eng. Units
+500.00
±
000.00
-500.00
10
µ
V
0C
±
150mV
Eng. Units
+150.00
±
000.00
-150.00
10
µ
V
0D
±
20mA
Eng. Units
+20.000
±
00.000
-20.000
1
µ
A
Code
Input Range
Data
Format
+Full
Scale
Zero
- Full
Scale
Disp.
Reso.
20
Pt-100, -100
°
C to
+100
°
C,
α
=0.00385
Eng. Units
+100.00
±
000.00
-100.00
0.01
°
C
21
Pt-100, 0
°
C to +100
°
C,
α
=0.00385
Eng. Units
+100.00
+000.00
+000.00
0.01
°
C
22
Pt-100, 0
°
C to +200
°
C,
α
=0.00385
Eng. Units
+200.00
+000.00
+000.00
0.01
°
C
23
Pt-100, 0
°
C to +600
°
C,
α
=0.00385
Eng. Units
+600.00
+000.00
+000.00
0.01
°
C
24
Pt-100, -100
°
C to
+100
°
C,
α
=0.003916
Eng. Units
+100.00
±
000.00
-100.00
0.01
°
C
25
Pt-100, 0
°
C to +100
°
C,
α
=0.003916
Eng. Units
+100.00
+000.00
+000.00
0.01
°
C