38
Program Temporary Basal Rate (TBR) continued
Step 1:
Put your insulin pump in
Run
mode.
Step 2:
Press the
Menu
key to move to the TEMPORARY BASAL RATE (TBR) screen.
Step 3:
Press the
Check
key to select.
Step 4:
Press or scroll the
Up
key to increase or the
Down
key to decrease the basal
rate in increments of 10% from 0% to 200%.
Step 5:
Press the
Menu
key to move to the TBR DURATION screen.
Step 6:
Press or scroll the
Up
key or
Down
key to correct and set the TBR DURATION.
Step 7:
Press the
Check
key to save and exit. Your Temporary Basal Rate will activate
immediately.
A TBR is displayed, which will remain for the duration you set in the TBR duration unless
you cancel this setting. When the TBR has ended, an A7: TBR OVER alert occurs. Press
the
CHECK
key twice to confi rm and turn off the alert.