![background image](http://html1.mh-extra.com/html/accu-chek/spirit/spirit_quick-reference-manual_3745958027.webp)
25
Cancel “Quick” Standard Bolus
During
programming
(bolus amount blinks):
• Press
the
Up
or
Down
key used in Step 1 on prior page to reset the bolus amount
to 0.0 units. The pump will play a melody. If you do not program a new bolus after
5 seconds, your insulin pump will return to RUN mode automatically and beep
three times.
No bolus is delivered
.
During
confi rmation
(beeps and vibrations occur), or during the
start delay
(Standard
bolus symbol blinks):
•
Press the
Up
key or
Down
key. The pump will play a melody. An A8: BOLUS
CANCELLED alert is displayed. Press the
Check
key twice to confi rm and turn off the
alert.
No bolus is delivered
.
During
bolus delivery
(countdown of bolus amount):
• Press and hold the
Up
key or
Down
key for 3 seconds until you hear a melody. An
A8: BOLUS CANCELLED alert is displayed. Press the
Check
key twice to confirm
and turn off the alert.
Bolus delivery is interrupted
.
The bolus amount delivered up to the cancellation can be reviewed in the
BOLUS HISTORY screen.