24
Program “Quick” Standard Bolus
Step 1:
From the RUN screen, press and hold the
Up
or
Down
key until you hear a
beep and feel a vibration.
Step 2:
Press the opposite
(Up
or
Down)
key used in Step 1 repeatedly until you see
the desired bolus amount. Your pump will simultaneously beep and vibrate
one time for each programmed bolus increment.
TIP:
Your pump comes programmed with a Standard bolus increment of 0.5. However,
bolus increments can be changed to 0.1, 0.2, 0.5, 1.0, and 2.0 units.
Five seconds after the last press of the
Up
or
Down
key, your pump will confi rm the total
bolus amount with one simultaneous beep and vibration for each bolus increment pro-
grammed, and the Standard bolus symbol will blink.
After a 5-second delay, your insulin pump will beep three times and start to deliver the
total bolus amount you programmed. The remaining bolus amount will count down on
the display.