55
Errors
Code Name
Meaning
Action
E11
Set Not
You changed the cartridge and • Press the
Check
key twice to turn
Primed
infusion set but did not prime
off and confi rm the error.
the infusion set.
• Remove or disconnect the infusion
set from your infusion site.
• Prime the infusion set.
• Reconnect the infusion set (tubing)
at your infusion site.
• Put your insulin pump into RUN
mode.
E12
Data
Data transfer between your
• Press the
Check
key twice to turn
Interrupted
insulin pump and PC or
off and confi rm the error.
Palm
®
device has been
• Restart the data transfer.
interrupted.
E13
Language Language choice needs to
• Press the
Check
key twice to turn
Error
be checked.
off and confi rm the error.
• Reset the language on your pump.
• Put your insulin pump into RUN mode.
Refer to ACCU-CHEK Spirit Insulin Pump User Guide for complete instruction, warnings, cautions and precautions.
CAUTIONS:
• No insulin is delivered when an Error Code appears on the display. Immediate
action must be taken to correct the Error Code.
• Never prime infusion set that is connected to your body.