Errors
Code Name
Meaning
Action
E8
Power
You either removed a battery
• Press the
Check
key twice to turn
Interrupt
without putting your pump into off and confi rm the error.
STOP fi rst, or you dropped
• Your insulin pump goes into
your pump.
STOP mode.
• Check the time and date, and
Note: If this error occurs
correct if needed.
during programming, check
• If a bolus and/or a Temporary Basal
your settings.
Rate was interrupted by the error,
review the Bolus History and/or the
Temporary Basal Rate History for
the delivered amount and duration.
• Put your insulin pump into RUN mode.
• If necessary, program a new bolus
and/or Temporary Basal Rate.
E10
Cartridge The
CHANGE THE CARTRIDGE • Press the
Check
key twice to turn
Error function was carried out
off and confi rm the error.
incorrectly.
•
Remove or disconnect the infusion
set from your infusion site.
• Move to the CHANGE THE CARTRIDGE
menu.
• Prime the infusion set.
• Reconnect the infusion set (tubing)
at your infusion site.
• Put you insulin pump into RUN mode.