![background image](http://html1.mh-extra.com/html/accu-chek/spirit/spirit_quick-reference-manual_3745958013.webp)
11
Insert Insulin Cartridge
Have the following materials ready:
• ACCU-CHEK Spirit insulin pump
• fi lled 3.15 ml plastic cartridge with an ACCU-CHEK adapter
• new ACCU-CHEK infusion set connected.
Step 1:
Put your insulin pump in STOP mode (select STOP YOUR ACCU-CHEK
screen and press
Check
key to confi rm setting).
Step 2:
Press the
Menu
key to move to the CHANGE THE CARTRIDGE screen.
Step 3:
Press the
Check
key to select.
Step 4:
Press and hold the
Check
key for 3 seconds until you hear a melody. The
piston rod will move backward.
CAUTION:
Do not touch or disturb the piston rod while it is rewinding. Doing so
may potentially damage your pump.
Step 5:
When the piston rod has returned completely, the pump will display the
CARTRIDGE VOLUME screen and the maximum cartridge volume (315 units).
Summary of Contents for Spirit