52
Errors
Code Name
Meaning
Action
E6
Mechanical
The
pump has detected a
• Press the
Check
key twice to turn
Error
mechanical problem.
off and confi rm the error.
• Remove or disconnect the infusion
Note: If this error continues,
set from your infusion site.
contact Disetronic
Pump
• Remove the cartridge, adapter,
Support
at
1-800-688-4578
infusion set (tubing), and battery
for more information. Contact
from your insulin pump. Insert a
your healthcare provider for an
new
battery.
alternate therapy plan.
• Rewind the piston rod.
• Install a new cartridge
and infusion set.
• Prime the infusion set.
• Reconnect the infusion set (tubing)
at your infusion site.
• Put you insulin pump into RUN mode.