10
Connect Insulin Cartridge, Adapter, and Infusion Set
Have the following materials ready:
• fi lled 3.15 ml plastic cartridge
• ACCU-CHEK
Spirit
adapter
• new ACCU-CHEK infusion set.
Step 1:
Remove the protective cap from the plastic cartridge.
CAUTION:
To prevent contamination, make sure that you do
not touch the plastic cartridge tip.
Step 2:
Push the adapter all the way onto the cartridge tip to the stop (a).
Step 3:
Hold the adapter and twist the infusion set luer-lock
connection by hand, clockwise into the adapter (b).
Step 4:
Carefully prepare a new ACCU-CHEK infusion set for use, priming and insertion,
following its instructions for use.
CAUTION:
Do not over-tighten.